Atomic Absorption Spectrophotometer (AAS)

Measures one metal at a time by how much light of that element's own wavelength is absorbed by free atoms in a flame or graphite furnace. Simple, robust and still widely used for metals in pharma QC.

  • Pharmaceutical
  • Food & beverage
  • Environmental
  • Mining & metallurgy
  • Agriculture
  • Clinical
Photo of a Atomic Absorption Spectrophotometer (AAS)
Photo: Talos (German Wikipedia), CC BY-SA 3.0 (cropped), via Wikimedia Commons
CALIBRATION

AAS calibration procedures

QUALIFICATION

Instrument qualification (DQ · IQ · OQ · PQ)

  1. 01

    DQ — Design qualification

    Defines needs such as flame and/or furnace, elements and detection limits required, background correction type and software with audit trail.

  2. 02

    IQ — Installation qualification

    Confirms correct delivery and installation, including gas supply, exhaust, electrical supply, lamps, software version and documentation.

  3. 03

    OQ — Operational qualification

    Tests wavelength accuracy, sensitivity (characteristic concentration), precision, baseline stability and background correction against specifications.

  4. 04

    PQ — Performance qualification

    Shows routine fitness for use, e.g. a sensitivity check standard and calibration curve meeting limits each day of use.