LimsPulse
Spectroscopy

UV-Visible Spectrophotometer

Measures how much ultraviolet or visible light a solution absorbs at each wavelength. Used in pharma QC for identification, assay, dissolution and many quick quantitative tests.

  • Pharmaceutical
  • Biotech
  • Food & beverage
  • Environmental
  • Chemicals
  • Academic research
Photo of a UV-Visible Spectrophotometer
Photo: Wiki210397, CC BY-SA 4.0 (cropped), via Wikimedia Commons
TROUBLESHOOTING

Common problems, causes and fixes

Noisy or drifting baseline
Likely causes
  • • Lamps not fully warmed up
  • • Deuterium lamp near end of life
  • • Dirty optics or cuvette windows
  • • Temperature changes in the room
What to do
  • • Allow a longer warm-up
  • • Check lamp energy and replace the lamp if low
  • • Clean the cuvettes; get optics cleaned by service
  • • Keep the instrument away from draughts and direct sunlight
Negative absorbance readings
Likely causes
  • • Blank absorbs more than the sample
  • • Baseline taken with a dirty or different cuvette
  • • Cuvettes not matched
What to do
  • • Prepare the blank from the same solvent lot
  • • Repeat the baseline with clean cuvettes
  • • Use a matched pair or always use the same cell
Results not reproducible
Likely causes
  • • Cuvette placed in different orientations
  • • Air bubbles or particles in the solution
  • • Fingerprints on the optical faces
What to do
  • • Mark the cuvette and always insert it the same way
  • • Tap out bubbles; filter or centrifuge turbid solutions
  • • Wipe the cuvette with lint-free tissue
Absorbance lower than expected at high concentration
Likely causes
  • • Stray light at high absorbance
  • • Solution outside the linear range
  • • Chemical effects such as association at high concentration
What to do
  • • Dilute so the reading falls in the linear range
  • • Check stray-light performance
  • • Use the dilution specified in the method
λmax shifted from the expected value
Likely causes
  • • Wavelength calibration error
  • • Wrong solvent or pH (many drugs shift with pH)
  • • Contaminated sample
What to do
  • • Run a wavelength accuracy check (holmium oxide or deuterium lines)
  • • Check the solvent and pH specified in the monograph
  • • Prepare a fresh solution
Very high or 'over-range' absorbance in the low UV
Likely causes
  • • Solvent cut-off (solvent absorbs strongly)
  • • Glass or plastic cuvette used in the UV
  • • Lamp failure
What to do
  • • Use a solvent with a suitable UV cut-off
  • • Use quartz cuvettes below about 320 nm
  • • Check lamp status and energy
In LimsPulse · Calibration Management

Keep every UV-Vis in control: usage logs, calibration and maintenance schedules, and automatic blocking of overdue instruments.

See how