Spectroscopy
Raman spectrometer
Shines a laser on a sample and measures the tiny fraction of light scattered with a shifted wavelength. Gives a sharp molecular fingerprint through glass and plastic, ideal for raw-material identity and polymorph checks.
- Pharmaceutical
- Biotech
- Forensics & security
- Materials science
- Semiconductors
- Art conservation

PREVENTIVE MAINTENANCE
Maintenance checklist
Daily / each use
- Wavenumber verification check
- Clean the probe window and nose cone
- Inspect fibre-optic cables (bench systems)
Monthly
- Trend verification results and laser power
- Check battery health on handheld units
- Review library for pending updates
Annual
- Calibration of wavenumber accuracy with an external certified standard
- Laser power and intensity (relative response) check
- Preventive maintenance by a trained engineer
SAFETY
Safety precautions
Before you work on the Raman
- Raman lasers are commonly Class 3B: never look into the beam or point it at people.
- Wear laser safety glasses rated for the laser wavelength when the beam path is open.
- Do not remove interlocked covers or defeat interlocks.
- Dark powders can heat or ignite under the laser; test with low power first.
- Follow the material's safety data sheet when handling raw materials.
In LimsPulse · Calibration Management
Keep every Raman in control: usage logs, calibration and maintenance schedules, and automatic blocking of overdue instruments.