Raman spectrometer
Shines a laser on a sample and measures the tiny fraction of light scattered with a shifted wavelength. Gives a sharp molecular fingerprint through glass and plastic, ideal for raw-material identity and polymorph checks.
- Pharmaceutical
- Biotech
- Forensics & security
- Materials science
- Semiconductors
- Art conservation

Raman calibration procedures
Instrument qualification (DQ · IQ · OQ · PQ)
- 01
DQ — Design qualification
Defines laser wavelength, spectral range, resolution, sampling accessories, library software and data integrity requirements.
- 02
IQ — Installation qualification
Confirms components, laser class labels, interlocks, software version and documentation.
- 03
OQ — Operational qualification
Checks wavenumber (x-axis) accuracy, laser wavelength, spectral resolution, intensity response and laser power with certified standards.
- 04
PQ — Performance qualification
Demonstrates correct identification of target materials and rejection of challenge materials with the validated library, plus routine daily checks.
Keep every Raman in control: usage logs, calibration and maintenance schedules, and automatic blocking of overdue instruments.