Polarimeter

Measures how much a solution of a chiral substance rotates the plane of polarised light. The result is used to calculate specific optical rotation, an identity and purity test for many APIs.

  • Pharmaceutical
  • Food & sugar
  • Flavours & fragrances
  • Chemicals
  • Academic research
Photo of a Polarimeter
Photo: Gingkoaceae, CC BY-SA 4.0 (cropped), via Wikimedia Commons
PREVENTIVE MAINTENANCE

Maintenance checklist

Daily / each use

  • Zero with the solvent blank
  • Check with a quartz control plate if your SOP requires it
  • Clean tubes and windows after use

Monthly

  • Inspect tube windows and seals for scratches or leaks
  • Clean the sample compartment
  • Check the temperature display against a calibrated probe (or as your SOP requires)

Six-monthly / annual

  • Calibrate optical rotation with certified quartz plates
  • Calibrate the sample temperature
  • Preventive maintenance and lamp check by the service engineer
  • Send quartz plates for recertification when due
SAFETY

Safety precautions

Before you work on the Polarimeter
  • Handle organic solvents in a fume hood and keep them away from ignition sources.
  • Glass tubes and windows break easily; handle them carefully and dispose of broken glass safely.
  • Do not look directly into an LED or lamp source with the cover open.
  • Wear gloves and eye protection when filling tubes with hazardous solutions.