LimsPulse
Thermal analysis & ovens

Melting Point Apparatus

Heats a sample packed in a thin glass capillary at a slow, controlled rate and records the temperatures at which it starts and finishes melting. Used for identity and purity checks of APIs and excipients.

  • Pharmaceutical
  • Fine chemicals
  • Agrochemicals
  • Dyes and pigments
  • Academic labs
Photo of a Melting Point Apparatus
Photo: Iain George, CC BY-SA 2.0 (cropped), via Wikimedia Commons
TROUBLESHOOTING

Common problems, causes and fixes

Melting points consistently too high
Likely causes
  • • Heating rate too fast
  • • Capillary too thick-walled or too wide
  • • Sensor out of calibration
What to do
  • • Use the specified ramp rate
  • • Use pharmacopoeial capillaries
  • • Calibrate with reference standards
Wide melting range
Likely causes
  • • Sample impure or not dried
  • • Powder too coarse or loosely packed
  • • Column too tall
What to do
  • • Dry the sample as per the monograph
  • • Grind finely and pack well
  • • Fill to about 3 mm only
Duplicates disagree
Likely causes
  • • Different packing in each capillary
  • • Capillaries inserted at different start temperatures
What to do
  • • Pack each capillary the same way
  • • Insert all capillaries at the same stabilised start temperature
Sample darkens or bubbles before melting
Likely causes
  • • Melting with decomposition
  • • Residual solvent or moisture
What to do
  • • Follow the monograph instructions for decomposing substances
  • • Dry the sample properly
Automatic detection gives a wrong point
Likely causes
  • • Dirty window or poor lighting
  • • Sample shrinks or sublimes before melting
What to do
  • • Clean the optics
  • • Review the video and confirm visually; set the evaluation threshold as validated
In LimsPulse · Calibration Management

Keep every Melting point apparatus in control: usage logs, calibration and maintenance schedules, and automatic blocking of overdue instruments.

See how