Laser Diffraction Particle Size Analyser
Measures the particle size distribution of powders, suspensions and emulsions from the pattern of laser light scattered by the particles. Reports D10, D50 and D90 within a minute or two.
- Pharmaceutical
- Biotech
- Chemicals
- Cosmetics
- Food & beverage
- Mining and cement

Laser diffraction PSA calibration procedures
Instrument qualification (DQ · IQ · OQ · PQ)
- 01
DQ — Design qualification
Defines the size range, dry and/or wet dispersion, software compliance (audit trail, electronic records) and throughput needed, and confirms the chosen model meets them.
- 02
IQ — Installation qualification
Verifies delivery, utilities (compressed air, vacuum, drain), software installation and documentation.
- 03
OQ — Operational qualification
Tests laser power, alignment, background and size accuracy and repeatability with a certified reference material.
- 04
PQ — Performance qualification
Shows the instrument gives consistent results on the lab's own methods, for example by repeat analysis of a control sample over time.
Keep every Laser diffraction PSA in control: usage logs, calibration and maintenance schedules, and automatic blocking of overdue instruments.