Microbiology & sterility
Laminar air flow unit (LAF)
A work station that blows HEPA-filtered air in a uniform, one-directional stream over the work surface to give an ISO 5 / Grade A zone for aseptic handling of media, samples and sterile materials.
- Pharmaceutical
- Biotech
- Hospitals and compounding pharmacies
- Electronics
- Food & beverage
- Research laboratories

TROUBLESHOOTING
Common problems, causes and fixes
Low air velocity
Likely causes
- • Blocked or loaded pre-filter or HEPA filter
- • Fan speed set too low or motor fault
- • Obstruction in front of the filter face
What to do
- • Replace the pre-filter and check differential pressure
- • Adjust fan speed and re-measure
- • Replace the HEPA filter if loaded, then re-qualify
High particle counts in the work zone
Likely causes
- • HEPA filter leak or damaged seal
- • Turbulence from room air or operator movement
- • Unit switched on too recently
- • Dirty work surface
What to do
- • Run a HEPA integrity (aerosol) test and repair or replace the filter
- • Check room airflow, close doors and reposition the unit
- • Allow the full stabilisation time
- • Clean the unit and repeat the count
Microbial growth on Grade A settle or contact plates
Likely causes
- • Poor aseptic technique
- • Inadequate cleaning or disinfectant contact time
- • Airflow disturbed or filter leak
- • Contaminated plates or materials
What to do
- • Raise a deviation and investigate as per the EM procedure
- • Identify the organism to find the likely source
- • Check air velocity and filter integrity
- • Review cleaning and operator training
Differential pressure gauge reading rising
Likely causes
- • HEPA or pre-filter loading with dust
What to do
- • Replace the pre-filter first
- • If the reading stays high and velocity drops, plan HEPA replacement
Differential pressure suddenly drops
Likely causes
- • Filter seal failure or torn filter
- • Fan failure or belt slip
What to do
- • Stop using the unit
- • Inspect the filter and fan; call service; re-qualify after repair
Alarm or unusual noise from the blower
Likely causes
- • Worn motor bearing
- • Loose panel or fan imbalance
What to do
- • Switch off and call maintenance
- • Do not use until the fault is fixed and velocity is re-checked