PREVENTIVE MAINTENANCE

Maintenance checklist

Daily / each use

  • Record background conductivity and pressure
  • Check eluent level or generator cartridge status
  • Check for leaks and salt deposits at fittings

Weekly

  • Replace eluent and rinse the reservoir
  • Check the autosampler and waste lines

Monthly / quarterly

  • Replace pump inlet filters
  • Check guard column performance and replace if peak shape degrades
  • Clean the suppressor if background rises

Six-monthly / annual

  • Preventive maintenance (pump seals, check valves)
  • Calibration of flow rate and conductivity detector
SAFETY

Safety precautions

Before you work on the IC
  • Hydroxide and acid eluents are corrosive: wear gloves and eye protection.
  • Eluent generators and electrolytic suppressors produce small amounts of hydrogen and oxygen gas; keep vents clear.
  • Never loosen fittings while the system is under pressure.
  • Dispose of eluent and sample waste as per site procedure.