Gel Permeation / Size-Exclusion Chromatography (GPC / SEC)

Separates molecules by their size in solution rather than by chemistry, so large molecules elute first. Used for polymer molecular-weight distribution and for protein aggregates in biologics.

  • Pharmaceutical
  • Biopharmaceutical
  • Polymers & plastics
  • Food (polysaccharides)
  • Cosmetics
  • Chemicals
Photo of a Gel Permeation / Size-Exclusion Chromatography (GPC / SEC)
Photo: TheBartgry, CC0 (cropped), via Wikimedia Commons
PREVENTIVE MAINTENANCE

Maintenance checklist

Daily / each use

  • Record pressure and RI baseline
  • Check solvent levels and waste
  • Check for leaks

Weekly

  • Prepare fresh aqueous mobile phase to prevent microbial growth
  • Check inlet filters

Monthly / quarterly

  • Check column performance with a reference standard (plate count, resolution of monomer and dimer)
  • Purge the RI reference cell

Six-monthly / annual

  • Preventive maintenance of pump and autosampler
  • Calibration of flow rate, oven temperature and detector
SAFETY

Safety precautions

Before you work on the GPC / SEC
  • THF is highly flammable and can form explosive peroxides; use stabilised THF, store it properly and work in a ventilated area.
  • Some GPC solvents (e.g. DMF, trichlorobenzene at high temperature) are toxic; follow the safety data sheet.
  • Never loosen fittings while the system is under pressure.
  • Dispose of solvent and biological waste as per site procedure.