GC Headspace Sampler
Heats a sealed vial containing the sample so volatile compounds move into the gas above it, then injects a portion of that gas into the GC. The standard way to test residual solvents without putting the sample matrix into the GC.
- Pharmaceutical
- Packaging
- Food & beverage
- Environmental
- Forensic
What is Headspace?
A static headspace sampler is an autosampler that sits on top of, or beside, a gas chromatograph. The sample is dissolved or suspended in a diluent (often water, DMSO, DMF or DMA) inside a crimp-sealed vial, typically 10 or 20 mL.
The vial is held at a set temperature for a set time (equilibration). Volatile solvents partition between the liquid and the gas above it (the headspace) until they reach equilibrium. The vial is then pressurised with carrier gas and a fixed volume of the headspace gas is transferred through a heated line into the GC inlet.
Because only gas is injected, non-volatile API and excipients stay in the vial. This keeps the inlet and column clean and makes headspace ideal for residual solvents testing to USP <467> and Ph. Eur. 2.4.24.
Used for
- Residual solvents in APIs, excipients and drug products
- Volatile impurities in packaging materials
- Ethanol content in liquids and blood-alcohol type analyses
- Monomer residues in polymers
- Volatile organic compounds in water
How it works
- 01
Prepare vial
Weigh the sample into a headspace vial, add diluent, and crimp the septum cap immediately.
- 02
Equilibrate
The vial is heated (e.g. 80 °C for 30–60 min), often with shaking, until solvents reach equilibrium with the headspace.
- 03
Pressurise
Carrier gas pressurises the vial to above the loop or column head pressure.
- 04
Fill loop
The vial vents through a sample loop (e.g. 1 mL), trapping a fixed volume of headspace gas.
- 05
Inject
A valve switches the loop into the carrier stream, which sweeps the gas through a heated transfer line into the GC.
- 06
Separate & detect
The GC separates the solvents and the FID measures each one.
Advantages
- Non-volatile matrix never enters the GC, so inlets and columns last longer
- Minimal sample preparation: weigh, add diluent, seal
- Good sensitivity for volatile solvents
- Fully automated with dozens of vials per sequence
Limitations
- Only for analytes volatile enough to enter the headspace at the chosen temperature
- Results depend strongly on matrix effects, so standards and samples need a matched diluent
- Precision is usually lower than direct liquid injection (%RSD limits are often wider)
- Long equilibration times limit throughput unless vials are overlapped
- Leaking vial seals give low, variable results