CALIBRATION

Colony counter calibration procedures

Colony counter has no instrument calibration of its own in most labs. Its performance is controlled through qualification (below) and the system suitability or acceptance checks of each test that uses it.

QUALIFICATION

Instrument qualification (DQ · IQ · OQ · PQ)

  1. 01

    DQ — Design qualification

    Defines manual or automated type, plate sizes, lighting, image storage and data integrity needs.

  2. 02

    IQ — Installation qualification

    Confirms the unit, software version, computer and network connections and user accounts are installed as specified.

  3. 03

    OQ — Operational qualification

    Checks lighting, magnification, counter function and, for automated systems, counting accuracy with reference plates or images, audit trail and security.

  4. 04

    PQ — Performance qualification

    Compares automated counts with manual counts by trained analysts on real plates across the countable range, showing agreement within the criterion set in the protocol.