CALIBRATION

CO₂ incubator calibration procedures

QUALIFICATION

Instrument qualification (DQ · IQ · OQ · PQ)

  1. 01

    DQ

    Define chamber volume, temperature and CO₂ range, sensor type, decontamination method, alarms and data logging.

  2. 02

    IQ

    Check installation site, CO₂ supply, regulators, filters, electrical connections and documentation.

  3. 03

    OQ

    Verify temperature accuracy and uniformity (empty-chamber mapping), CO₂ accuracy, recovery after door opening, and alarms.

  4. 04

    PQ

    Repeat temperature and CO₂ checks under typical load and confirm cultures grow as expected over a defined period.

In LimsPulse · Calibration Management

Keep every CO₂ incubator in control: usage logs, calibration and maintenance schedules, and automatic blocking of overdue instruments.