LimsPulse
Bioanalytical

CO₂ incubator

A warm, humid chamber that holds mammalian cell cultures at body temperature (usually 37 °C), high humidity and a set CO₂ level (usually 5%), so the culture medium stays at the right pH.

  • Biotechnology
  • Biopharmaceutical
  • Cell and gene therapy
  • IVF clinics
  • Academic research
  • Vaccine manufacture
Photo of a CO₂ incubator
Photo: Unknown (public domain), Public domain (cropped), via Wikimedia Commons
Follow your site SOP
This is a typical operating sequence for learning. Your instrument model and approved SOP take precedence.
STAGE 1

Before you start

  1. 1Check the incubator is within its verification due date and the display shows set temperature and CO₂.
  2. 2Check CO₂ cylinder pressure and that the auto-changeover (if fitted) is set.
  3. 3Check the water pan is filled with sterile purified water.
  • Secure CO₂ cylinders upright with a chain or strap.
STAGE 2

Use

  1. 1Disinfect gloves and vessel surfaces before placing cultures inside.
  2. 2Place flasks with vented caps (or caps loosened) so gas can exchange.
  3. 3Open the door as briefly as possible and not more often than needed.
  4. 4Record entries and any alarms in the logbook or monitoring system.
  • Do not overload shelves; air must circulate around vessels.
STAGE 3

After use / end of day

  1. 1Wipe spills at once with a suitable disinfectant, then sterile water.
  2. 2Check the display values have recovered after the last opening.
In LimsPulse · Calibration Management

Keep every CO₂ incubator in control: usage logs, calibration and maintenance schedules, and automatic blocking of overdue instruments.