Analytical balance

Measures the mass of a sample very precisely, usually to 0.1 mg or 0.01 mg. Almost every quantitative test in a QC lab starts with a weighing, so the balance is the most used instrument in the lab.

  • Pharmaceutical
  • Biotech
  • Chemicals
  • Food & beverage
  • Academic research
  • Cosmetics
Photo of a Analytical balance
Photo: Karelj, Public domain (cropped), via Wikimedia Commons
PREVENTIVE MAINTENANCE

Maintenance checklist

Daily / each use

  • Check levelling
  • Clean the pan and chamber
  • Perform and record the daily performance check
  • Check the internal adjustment has run, if the balance has one

Weekly

  • Clean the draught shield glass and housing with a lint-free cloth
  • Check the anti-vibration table and surroundings for new sources of draught or vibration

Monthly

  • Repeatability check (or as defined in your SOP)
  • Check the test weights are clean, in their box and within their calibration date

Six-monthly / annual

  • Full calibration (accuracy, repeatability, eccentricity, linearity, minimum weight) by a trained person or an accredited service provider
  • Recalibration of the test weights by an accredited laboratory
SAFETY

Safety precautions

Before you work on the Analytical balance
  • Weigh potent, toxic or sensitising powders in a ventilated balance enclosure, wearing gloves and, where required, a mask.
  • Clean spills immediately using the approved cleaning procedure; do not blow powder away.
  • Handle test weights only with forceps or gloves to avoid contaminating them.
  • Do not lean on the balance bench or place heavy objects on it while weighing.