Follow your site SOP
This is a typical operating sequence for learning. Your instrument model and approved SOP take precedence.
STAGE 1
Before loading samples
- 1Check the chamber is qualified, calibrated and within due dates, and that the monitoring system shows it is stable at the set conditions.
- 2Confirm the study protocol: condition, number of samples, pull points and storage location.
- 3Check the humidifier water reservoir is filled with purified water.
- Never place samples in a chamber that is in alarm or has an open deviation.
STAGE 2
Loading
- 1Label every sample with study number, condition, pull point and date.
- 2Place samples only in the qualified (mapped) zone, leaving space between packs and walls for air to flow.
- 3Record the location (shelf and position) in the stability log or LIMS.
- 4Close the door quickly and check that conditions recover within the time established during qualification.
- Do not block fans or air inlets. Overloading creates hot and wet spots that were not covered by mapping.
STAGE 3
Routine use and sample pulls
- 1Review the monitoring record daily (or as per SOP) and sign it.
- 2Pull samples on the scheduled date, within the window allowed by your SOP, and record the pull.
- 3Keep door openings short and plan pulls together to limit disturbance.
- Any excursion must be recorded and assessed, even if conditions recovered quickly.
STAGE 4
Alarm or failure
- 1Respond to the alarm, check the cause (door left open, water empty, power, compressor fault).
- 2If the fault cannot be fixed quickly, move samples to a qualified backup chamber at the same condition.
- 3Raise a deviation and assess the impact of the excursion on every study in the chamber.