Stability chamber

A cabinet or walk-in room that holds samples at a controlled temperature and relative humidity for months or years, so that the shelf life of a drug product can be established under ICH storage conditions.

  • Pharmaceutical
  • Biotech
  • Medical devices
  • Cosmetics
  • Food & nutraceuticals
  • Packaging
Photo of a Stability chamber
Photo: Encik Tekateki, CC BY-SA 4.0 (cropped), via Wikimedia Commons
PREVENTIVE MAINTENANCE

Maintenance checklist

Daily

  • Review temperature and RH records and alarms
  • Check the water reservoir level
  • Check the door closes and seals properly

Monthly

  • Clean the interior and shelves
  • Drain and clean the humidifier reservoir to prevent scale and microbial growth
  • Check the condensate drain is not blocked

Six-monthly

  • Calibrate control and monitoring sensors
  • Test high and low alarms for temperature and RH
  • Clean condenser coils and check fan operation

Annual / after change

  • Preventive maintenance by a trained engineer (refrigerant, humidifier, door gaskets)
  • Temperature and RH mapping, typically annually or after relocation, major repair or a change in load
SAFETY

Safety precautions

Before you work on the Stability chamber
  • Walk-in chambers must have an inside door release and an alarm so nobody can be trapped.
  • Hot humid air at 40 °C/75% RH can cause discomfort; limit time spent inside walk-in rooms.
  • Isolate power before cleaning heaters, fans or the humidifier.
  • Use only purified water in the humidifier to prevent scale and microbial growth.