LimsPulse
Cosmetics testing

In vitro SPF analyser (UV transmittance analyser)

Measures how much ultraviolet light passes through a thin film of sunscreen spread on a roughened PMMA plate, then calculates in vitro SPF, UVA protection factor and critical wavelength from the transmittance spectrum.

  • Cosmetics & personal care
  • Sunscreen manufacturers
  • UV filter suppliers
  • Contract testing labs
  • Textiles (UV protection)
Illustration of In vitro SPF analyser (UV transmittance analyser)
TROUBLESHOOTING

Common problems, causes and fixes

Large plate-to-plate variation
Likely causes
  • • Uneven spreading
  • • Wrong product dose
  • • Plates of different roughness or batch
  • • Film not settled
What to do
  • • Train analysts and standardise spreading time and pressure
  • • Weigh the dose on a verified balance
  • • Use one plate lot per study and check roughness certificates
  • • Keep to the drying time
In vitro SPF far above in vivo value
Likely causes
  • • Too much product applied
  • • Plate roughness too low
  • • Fluorescence from the product
What to do
  • • Check the application dose and balance
  • • Use plates that meet the method roughness specification
  • • Check the instrument's fluorescence handling or use filters
Noisy spectrum below 300 nm
Likely causes
  • • Low lamp output at short wavelengths
  • • Very high absorbance (little light passes)
  • • Dirty optics
What to do
  • • Check lamp hours and replace if needed
  • • Confirm the dynamic range of the instrument covers the absorbance measured
  • • Clean windows and sphere port
Reference sunscreen out of range
Likely causes
  • • Instrument drift
  • • Plate or glycerine blank problem
  • • Expired or badly stored reference product
What to do
  • • Repeat the blank and re-verify transmittance with filters
  • • Use fresh plates and blank
  • • Replace the reference product
Wavelength check fails
Likely causes
  • • Monochromator misalignment
  • • Filter damaged or out of certification
What to do
  • • Call service to realign
  • • Recertify or replace the filter
In LimsPulse · Calibration Management

Keep every SPF analyser in control: usage logs, calibration and maintenance schedules, and automatic blocking of overdue instruments.

See how