Cosmetics testing
In vitro SPF analyser (UV transmittance analyser)
Measures how much ultraviolet light passes through a thin film of sunscreen spread on a roughened PMMA plate, then calculates in vitro SPF, UVA protection factor and critical wavelength from the transmittance spectrum.
- Cosmetics & personal care
- Sunscreen manufacturers
- UV filter suppliers
- Contract testing labs
- Textiles (UV protection)
PREVENTIVE MAINTENANCE
Maintenance checklist
Daily / each use
- Check lamp warm-up and baseline stability
- Clean the sample stage and optical windows with lint-free tissue
- Store PMMA plates clean, dry and away from dust
Monthly
- Run the reference sunscreen and trend its result
- Check lamp hours and intensity against the manufacturer's limit
- Back up spectra and results
Every 6–12 months
- Wavelength and transmittance verification with certified filters
- Solar simulator irradiance check with a calibrated radiometer
- Preventive maintenance by the supplier
As needed
- Replace the lamp when intensity or noise is out of specification
- Re-verify after lamp change, repair or relocation
SAFETY
Safety precautions
Before you work on the SPF analyser
- UV radiation damages eyes and skin. Keep covers closed and use UV-blocking glasses near open sources.
- Xenon lamps run hot and are under pressure; let them cool before changing and follow the manual.
- Some UV filters and solvents are skin sensitisers; wear gloves when spreading products.
- Dispose of used plates and wipes as chemical waste if the products contain solvents.