In vitro SPF analyser (UV transmittance analyser)
Measures how much ultraviolet light passes through a thin film of sunscreen spread on a roughened PMMA plate, then calculates in vitro SPF, UVA protection factor and critical wavelength from the transmittance spectrum.
- Cosmetics & personal care
- Sunscreen manufacturers
- UV filter suppliers
- Contract testing labs
- Textiles (UV protection)
Follow your site SOP
This is a typical operating sequence for learning. Your instrument model and approved SOP take precedence.
STAGE 1
Before you start
- 1Switch on the lamp and allow the warm-up time in the manual (often 15–30 minutes).
- 2Check the wavelength and transmittance verification is in date.
- 3Condition PMMA plates and products at room temperature (about 25 °C).
- 4Confirm the analytical balance used for film dosing is verified.
- UV lamps and solar simulators can burn skin and damage eyes. Never look into the beam and keep covers closed.
STAGE 2
Reference and sample plates
- 1Measure a blank reference plate (glycerine or as the method states) to set the 100% transmittance baseline.
- 2Weigh the product onto a fresh plate as many small dots at the specified dose (for ISO 24443 typically about 1.3 mg/cm²).
- 3Spread quickly and evenly over the whole plate, first with light strokes and then with firmer pressure, using a pre-saturated gloved finger or an automated spreader.
- 4Let the film settle in the dark for the time in the method.
- Keep spreading time and pressure consistent. Most variation in in vitro SPF comes from the film, not the instrument.
STAGE 3
Measurement
- 1Scan each plate at several positions (for example 4–9 spots) from 290 to 400 nm.
- 2Test at least three plates per product (or the number in the method).
- 3For UVA-PF, expose the plates to the calculated UV dose in the solar simulator and scan again.
- 4Run a reference sunscreen of known value with each series as a control.
- Reject spots with visible defects, bubbles or bare patches as the method allows, and record why.
STAGE 4
Results and shutdown
- 1Review the spectra for unusual shapes, then calculate SPF, UVA-PF, UVA-PF/SPF ratio and critical wavelength.
- 2Check variability between plates against the method limit (for example a coefficient of variation or 95% confidence interval).
- 3Discard used plates (single use), clean the stage and switch off the lamp.
- 4Complete the instrument log.