In vitro SPF analyser (UV transmittance analyser)
Measures how much ultraviolet light passes through a thin film of sunscreen spread on a roughened PMMA plate, then calculates in vitro SPF, UVA protection factor and critical wavelength from the transmittance spectrum.
- Cosmetics & personal care
- Sunscreen manufacturers
- UV filter suppliers
- Contract testing labs
- Textiles (UV protection)
SPF analyser calibration procedures
Instrument qualification (DQ · IQ · OQ · PQ)
- 01
DQ — Design qualification
Defines needs: wavelength range 290–400 nm, resolution, dynamic range of absorbance, integrating sphere, solar simulator, plate types and software compliant with data-integrity rules.
- 02
IQ — Installation qualification
Confirms the instrument, solar simulator, software and accessories are installed as specified, with serial numbers, manuals and environmental conditions recorded.
- 03
OQ — Operational qualification
Verifies wavelength accuracy, transmittance accuracy and linearity with certified filters, baseline stability, stray light and solar simulator irradiance.
- 04
PQ — Performance qualification
Shows the system gives correct results in routine use, for example a reference sunscreen measured on several plates falls within its expected range.
Keep every SPF analyser in control: usage logs, calibration and maintenance schedules, and automatic blocking of overdue instruments.