Particle & solid-state
Sieve Shaker (Analytical Sieving)
Shakes a stack of test sieves so a powder or granule separates into size fractions. Weighing each sieve gives the particle size distribution by mass.
- Pharmaceutical
- Food & beverage
- Chemicals
- Mining and construction
- Agriculture

TROUBLESHOOTING
Common problems, causes and fixes
Mass balance poor (loss above limit)
Likely causes
- • Dust escaping
- • Material stuck in sieves
- • Weighing errors
What to do
- • Fit the lid and seals properly
- • Brush out sieves and re-weigh
- • Check balance and tare values
Results coarser than expected
Likely causes
- • Sieve blinding
- • Sieving time too short
- • Agglomerated or electrostatically charged powder
What to do
- • Clean sieves; use sieving aids if validated
- • Confirm the end-point
- • Condition the sample or use air-jet sieving
Results finer than expected
Likely causes
- • Torn or stretched mesh
- • Granule attrition during shaking
What to do
- • Inspect and replace damaged sieves
- • Reduce amplitude or time within validated limits
Shaker noisy or stack moving
Likely causes
- • Loose clamp
- • Worn rubber mounts
What to do
- • Tighten the clamp
- • Replace mounts