Sieve Shaker (Analytical Sieving)

Shakes a stack of test sieves so a powder or granule separates into size fractions. Weighing each sieve gives the particle size distribution by mass.

  • Pharmaceutical
  • Food & beverage
  • Chemicals
  • Mining and construction
  • Agriculture
Photo of a Sieve Shaker (Analytical Sieving)
Photo: Cjp24, CC BY-SA 3.0 (cropped), via Wikimedia Commons
TROUBLESHOOTING

Common problems, causes and fixes

Mass balance poor (loss above limit)
Likely causes
  • • Dust escaping
  • • Material stuck in sieves
  • • Weighing errors
What to do
  • • Fit the lid and seals properly
  • • Brush out sieves and re-weigh
  • • Check balance and tare values
Results coarser than expected
Likely causes
  • • Sieve blinding
  • • Sieving time too short
  • • Agglomerated or electrostatically charged powder
What to do
  • • Clean sieves; use sieving aids if validated
  • • Confirm the end-point
  • • Condition the sample or use air-jet sieving
Results finer than expected
Likely causes
  • • Torn or stretched mesh
  • • Granule attrition during shaking
What to do
  • • Inspect and replace damaged sieves
  • • Reduce amplitude or time within validated limits
Shaker noisy or stack moving
Likely causes
  • • Loose clamp
  • • Worn rubber mounts
What to do
  • • Tighten the clamp
  • • Replace mounts