Refrigerator / deep freezer

Pharmaceutical-grade cold storage for reference standards, reagents, culture media, samples and temperature-sensitive products, typically at 2–8 °C (refrigerator), −20 °C (freezer) or −80 °C (ultra-low).

  • Pharmaceutical
  • Biotech
  • Hospitals & blood banks
  • Clinical research
  • Food testing
Photo of a Refrigerator / deep freezer
Photo: Magnus Manske, CC BY-SA 3.0 (cropped), via Wikimedia Commons
PREVENTIVE MAINTENANCE

Maintenance checklist

Daily

  • Record or review temperatures
  • Check the door seal and that the door closes fully

Monthly

  • Clean shelves and interior; remove expired material
  • Check ice build-up in freezers
  • Test the alarm (or as per SOP)

Quarterly / six-monthly

  • Clean condenser coils and filters
  • Defrost manual-defrost freezers
  • Calibrate the display probe and data logger

Annual / after change

  • Preventive maintenance by a trained engineer
  • Temperature mapping, typically annually or after relocation, repair or change in use
SAFETY

Safety precautions

Before you work on the Fridge / freezer
  • Only spark-free (flammable-safe) units may hold flammable solvents.
  • Use insulated gloves and eye protection with −80 °C freezers and dry ice.
  • Never store food or drink with lab materials.
  • Keep cold rooms fitted with inside door releases and alarms.