Membrane filtration unit

A manifold with filter funnels connected to a vacuum source (or a peristaltic pump for closed canisters) that passes liquid samples through a 0.45 µm membrane so that microorganisms are retained and can be grown and counted.

  • Pharmaceutical
  • Biotech
  • Water testing
  • Food & beverage
  • Medical devices
Photo of a Membrane filtration unit
Photo: Stéphane Lesbats (Ifremer), CC BY 4.0 (cropped), via Wikimedia Commons
TROUBLESHOOTING

Common problems, causes and fixes

Filtration very slow or stops
Likely causes
  • • Sample with particles or high viscosity
  • • Weak vacuum or leak
What to do
  • • Pre-dilute or use several membranes
  • • Check seals and pump
Growth on negative controls
Likely causes
  • • Non-sterile funnels or rinse fluid
  • • Poor aseptic technique
  • • Contaminated environment
What to do
  • • Check sterilisation records
  • • Retrain analyst and review EM data
  • • Repeat with fresh materials
Low recovery in method suitability
Likely causes
  • • Residual antimicrobial on the membrane
  • • Wrong membrane type
What to do
  • • Increase rinse volume or add neutraliser
  • • Try another membrane material
Liquid in vacuum line
Likely causes
  • • Trap full
  • • Hydrophobic filter missing or wet
What to do
  • • Empty trap and replace filter
  • • Service the pump