Kinematic Viscosity Bath

A clear, stirred liquid bath held to within about ±0.02 °C, in which calibrated glass capillary viscometers are timed. Flow time multiplied by the viscometer constant gives kinematic viscosity in mm²/s (cSt).

  • Lubricant blending
  • Petroleum refining
  • Fuel terminals
  • Marine fuel testing
  • Fleet and machinery oil condition monitoring
  • Independent testing laboratories
Illustration of Kinematic Viscosity Bath
TROUBLESHOOTING

Common problems, causes and fixes

Two flow times disagree
Likely causes
  • • Sample not at temperature yet
  • • Bubble or particle in the capillary
  • • Viscometer not vertical
  • • Bath temperature cycling
What to do
  • • Allow longer equilibration
  • • Filter sample and refill
  • • Check holder alignment
  • • Tune the controller or improve stirring
Results biased high or low on a check standard
Likely causes
  • • Bath temperature offset
  • • Wrong viscometer constant entered
  • • Residual solvent or water in viscometer
What to do
  • • Check temperature with a calibrated thermometer
  • • Check constant against the certificate for the test temperature
  • • Clean and dry thoroughly
Flow time below 200 s
Likely causes
  • • Viscometer too large for the sample
What to do
  • • Use a smaller-capillary viscometer
Meniscus hard to see
Likely causes
  • • Dark or opaque sample
  • • Cloudy bath medium
What to do
  • • Use a reverse-flow (opaque) viscometer
  • • Replace bath medium and clean the glass
Bath cannot reach or hold temperature
Likely causes
  • • Lid missing, heat loss high
  • • Heater or chiller fault
  • • Medium level low
What to do
  • • Fit the lid
  • • Service heater/chiller
  • • Top up medium