Franz Diffusion Cell (Vertical Diffusion Cell)

A two-chamber glass cell used to measure how fast drug is released from creams, gels and ointments, or permeates through skin. Drug passes through a membrane into a stirred receptor medium that is sampled over time.

  • Pharmaceutical (topicals and transdermals)
  • Cosmetics
  • Research and academia
Photo of a Franz Diffusion Cell (Vertical Diffusion Cell)
Photo: RuiCiencia, CC BY 4.0 (cropped), via Wikimedia Commons
TROUBLESHOOTING

Common problems, causes and fixes

Low, variable release in some cells
Likely causes
  • • Air bubble under the membrane
  • • Uneven dosing
  • • Membrane wrinkled
What to do
  • • Remove bubbles before dosing
  • • Weigh the dose and spread evenly
  • • Mount membrane flat
Receptor medium leaking
Likely causes
  • • Clamp loose
  • • Chipped joint
What to do
  • • Tighten clamp
  • • Replace damaged cell
Release levels off early
Likely causes
  • • Loss of sink conditions
  • • Product drying out
What to do
  • • Use a larger receptor volume or add solubiliser
  • • Occlude the donor chamber
Temperature out of range
Likely causes
  • • Water bath set wrongly
  • • Poor circulation
What to do
  • • Adjust bath to give 32 °C at the membrane
  • • Check circulation tubing