Franz Diffusion Cell (Vertical Diffusion Cell)

A two-chamber glass cell used to measure how fast drug is released from creams, gels and ointments, or permeates through skin. Drug passes through a membrane into a stirred receptor medium that is sampled over time.

  • Pharmaceutical (topicals and transdermals)
  • Cosmetics
  • Research and academia
Photo of a Franz Diffusion Cell (Vertical Diffusion Cell)
Photo: RuiCiencia, CC BY 4.0 (cropped), via Wikimedia Commons
Follow your site SOP
This is a typical operating sequence for learning. Your instrument model and approved SOP take precedence.
STAGE 1

Setup

  1. 1Check cells are clean and their receptor volumes and orifice diameters are recorded.
  2. 2Degas the receptor medium and fill each receptor chamber, tilting to remove bubbles.
  3. 3Mount the pre-soaked membrane and clamp the donor chamber in place.
  4. 4Allow the cells to equilibrate to temperature with stirring.
  • Check under each membrane for bubbles; tilt the cell to remove them before dosing.
STAGE 2

Dosing and sampling

  1. 1Apply the stated amount of product evenly over the membrane and start timing.
  2. 2At each time point withdraw the stated volume from the sampling arm.
  3. 3Immediately replace with the same volume of fresh, warmed medium.
  4. 4Run the stated number of cells (commonly six) for each product.
  • Keep sampling times exact and stagger cells so each is sampled on time.
STAGE 3

After the test

  1. 1Analyse samples with a validated method (usually HPLC).
  2. 2Dismantle and clean cells; discard membranes.
  3. 3Calculate cumulative release correcting for amounts removed at earlier samples.