Centrifuge
Spins tubes at high speed so that the centrifugal force separates solids from liquids or liquids of different density. In QC it is used to clarify sample extracts and separate phases.
- Pharmaceutical
- Biotech
- Clinical laboratories
- Food & dairy
- Research

TROUBLESHOOTING
Common problems, causes and fixes
Imbalance alarm or vibration
Likely causes
- • Unequal tube masses
- • Tubes not placed opposite each other
- • Missing bucket
What to do
- • Re-balance tubes by mass
- • Load symmetrically
- • Fit all buckets, even if empty
Supernatant still cloudy
Likely causes
- • RCF too low
- • Time too short
- • Fine colloidal particles
What to do
- • Check the method RCF and radius
- • Increase time within method limits
- • Use a higher-speed rotor or filtration as allowed
Lid will not open
Likely causes
- • Rotor still turning
- • Power failure
What to do
- • Wait for the rotor to stop
- • Use the manual emergency release only after confirming the rotor has stopped
Does not reach set speed
Likely causes
- • Worn motor brushes or drive fault
- • Rotor not recognised
What to do
- • Call service
- • Check rotor seating and identification
Tubes cracked
Likely causes
- • Tube RCF rating exceeded
- • Old or solvent-damaged tubes
What to do
- • Use tubes rated for the RCF
- • Use solvent-compatible tubes and replace old stock