TROUBLESHOOTING

Common problems, causes and fixes

Drops cling to the inside wall
Likely causes
  • • Grease or dirt on the glass
What to do
  • • Clean with laboratory detergent or a suitable cleaning solution and rinse well
Burette reading drifts with stopcock closed
Likely causes
  • • Leaking stopcock
What to do
  • • Tighten or replace the stopcock
Air bubble in the tip changes during titration
Likely causes
  • • Tip not cleared before setting zero
What to do
  • • Run titrant through the tip before setting zero
Gravimetric check fails
Likely causes
  • • Chipped tip
  • • Wrong delivery or waiting time
  • • Temperature not recorded or wrong Z factor
What to do
  • • Withdraw chipped glassware
  • • Follow the delivery time marked
  • • Measure water temperature and use the correct Z factor