Blaine Air Permeability Apparatus

Measures the fineness of cement as specific surface area (m²/kg) by timing how long a fixed volume of air takes to pass through a compacted bed of cement.

  • Cement manufacturing
  • Construction materials testing
  • Fly ash and slag suppliers
  • Academic labs
Illustration of Blaine Air Permeability Apparatus
TROUBLESHOOTING

Common problems, causes and fixes

Flow times much longer than expected
Likely causes
  • • Bed over-compacted or too much sample
  • • Blocked perforated disc
  • • Wrong filter paper
What to do
  • • Recalculate and reweigh the sample
  • • Clean the disc
  • • Use the specified filter paper
Flow times very short
Likely causes
  • • Air leak at the seal
  • • Too little sample
  • • Bed cracked or channelled
What to do
  • • Reseat or replace the seal
  • • Check the calculation
  • • Re-form the bed
Results drift over months on a control cement
Likely causes
  • • Manometer liquid evaporated or contaminated
  • • Plunger worn
  • • Temperature differs from calibration
What to do
  • • Refill to the mark
  • • Re-determine the constant
  • • Apply a temperature correction or control room temperature
Duplicate times differ widely
Likely causes
  • • Inconsistent bed preparation
  • • Operator timing error
What to do
  • • Standardise compaction technique
  • • Use an automatic timing unit