Detector wavelength accuracy

Confirms the UV detector reads the correct wavelength, using a compound with known absorbance maxima or the built-in holmium oxide filter.

Photo of a HPLC
Photo: Fernando Frick, CC0 (cropped), via Wikimedia Commons
PURPOSE

Why this is checked

If the detector wavelength is off, sensitivity drops and results change, especially for methods measured on the slope of an absorbance band.

Frequency

Typically every 6 months, and after major repair or relocation, as defined in your calibration SOP.

Learn how the HPLC works
YOU WILL NEED

Standards & tools

  • Caffeine solution in water (maxima near 205 nm and 273 nm)
  • or the detector's built-in holmium oxide filter (lines near 361, 418 and 536 nm)
BEFORE YOU START

Prerequisites

  • Allow the lamp to warm up fully
  • Flush the flow cell with water
PROCEDURE

Step by step

  1. 1Fill the flow cell with the caffeine solution (or start the built-in holmium oxide test).
  2. 2Scan or step the wavelength across each maximum (e.g. 265–281 nm in 1 nm steps).
  3. 3Record the wavelength giving the highest absorbance for each maximum.
  4. 4Compare the observed maxima with the reference values.
RECORD

Example observation table

Reference maximum (nm)Observed maximum (nm)Difference (nm)
205206+1
2732730
CALCULATION

How to calculate the result

Difference = Observed maximum − Reference maximum
  1. Reference 273 nm, observed 273 nm
  2. Difference = 0 nm (within ±2 nm)
ACCEPTANCE CRITERIA

Pass if

  • Observed maxima within ±2 nm of the reference value (tighter limits may apply to PDA detectors).
IF IT FAILS

What to do

  • Check lamp energy and age; replace the lamp if low.
  • Recalibrate the wavelength using the manufacturer's procedure.
  • Do not use the detector until it passes.