HPLC · Calibration procedure

Column oven temperature accuracy

Verifies the column oven holds the set temperature, using a calibrated temperature probe.

Photo of a HPLC
Photo: Fernando Frick, CC0 (cropped), via Wikimedia Commons
PURPOSE

Why this is checked

Retention time and selectivity change with temperature. A drifting oven makes retention times and resolution unreliable.

Frequency

Typically every 6 months, and after major repair or relocation, as defined in your calibration SOP.

Learn how the HPLC works
YOU WILL NEED

Standards & tools

  • Calibrated digital thermometer with a thin probe, traceable to national standards
BEFORE YOU START

Prerequisites

  • Place the probe in the column compartment at the column position
  • Close the oven door and let it stabilise
PROCEDURE

Step by step

  1. 1Set the oven to a low temperature (e.g. 30 °C) and wait at least 15 minutes after it reaches the set point.
  2. 2Record the probe reading.
  3. 3Repeat at a mid and a high temperature used by your methods (e.g. 40 °C and 60 °C).
RECORD

Example observation table

Set (°C)Observed (°C)Difference (°C)
3030.4+0.4
4039.8−0.2
6059.3−0.7
CALCULATION

How to calculate the result

Difference = Observed − Set
  1. Set 60 °C, observed 59.3 °C
  2. Difference = −0.7 °C (within ±2 °C)
ACCEPTANCE CRITERIA

Pass if

  • Observed temperature within ±2 °C of the set value.
IF IT FAILS

What to do

  • Check the door seal and probe position
  • Service the oven heater or sensor
  • Repeat after repair