Measures temperature (thermometer) or records it over time (data logger). Used to check water baths, ovens, incubators, fridges, stability chambers and sample temperatures, and to map storage areas.

  • Pharmaceutical
  • Food
  • Logistics
  • Clinical labs
  • Biotech
Photo of a Thermometer & Temperature Data Logger
Photo: Aperegoy, CC BY-SA 4.0 (cropped), via Wikimedia Commons
LEARN & PRACTICE

Test your Thermometer / logger knowledge

Multiple-choice questions

Pick an answer to see whether you are right and why.

  1. Q1. What resistance does a Pt100 sensor have at 0 °C?

    basic
  2. Q2. What makes a reference thermometer traceable?

    basic
  3. Q3. Reference reads 5.02 °C with a certificate correction of +0.03 °C; the logger reads 5.3 °C. What is the logger error?

    intermediate
  4. Q4. How should calibration test points be chosen for a logger?

    intermediate
  5. Q5. Which sensor is generally most stable and accurate, and so often used as a reference?

    basic

Viva questions

Think of your answer first, then open it to compare.

  1. 1. Describe a comparison calibration of a data logger.

    Place the logger sensor and a calibrated reference thermometer side by side in a stable stirred bath or dry block at each test point. When stable, record both readings, apply the reference correction, calculate the logger error at each point and compare with the SOP limit.

  2. 2. A logger in a stability chamber shows a gap in its data. What do you do?

    Investigate the cause (battery, memory, start settings), check other monitoring records for the gap period, assess impact on the stored samples, record a deviation as required and correct the cause, for example by replacing the battery and reviewing the logger schedule.