A very sensitive balance that reads to 0.001 mg (1 µg) or better. It is used when the sample is only a few milligrams, for example a potent reference standard or a DSC sample.

  • Pharmaceutical
  • Biotech
  • Chemicals
  • Environmental testing
  • Academic research
  • Materials science
TROUBLESHOOTING

Common problems, causes and fixes

Reading drifts up or down by many micrograms
Likely causes
  • • Static charge on the container or sample
  • • Sample or container not at room temperature
  • • Draught from air-conditioning or an open door
What to do
  • • Use the ioniser and anti-static containers
  • • Allow more time to equilibrate; handle with forceps
  • • Close doors, redirect air flow, use a smaller draught shield
Stability symbol never appears
Likely causes
  • • Vibration from nearby equipment or people walking
  • • Balance not level
  • • Weighing table touching the bench or wall
What to do
  • • Move vibrating equipment away; weigh at quieter times
  • • Level the balance
  • • Make sure the stone table stands free
Zero is not stable after removing the sample
Likely causes
  • • Material left on the pan
  • • Temperature changed since the last tare
  • • Pan touching the draught shield
What to do
  • • Clean the pan and re-zero
  • • Wait for stabilisation, then re-zero
  • • Re-seat the pan correctly
Repeatability or minimum weight check fails
Likely causes
  • • Poor environment at the place of use
  • • Weight handled with fingers
  • • Worn or dirty weighing cell parts
What to do
  • • Improve the environment and repeat
  • • Use forceps and a clean, calibrated weight
  • • Call service; raise the working minimum weight until fixed
Reading slowly increases for a hygroscopic sample
Likely causes
  • • Sample absorbs moisture from the air
What to do
  • • Weigh quickly in a closed container
  • • Weigh by difference
  • • Control the room humidity