A very sensitive balance that reads to 0.001 mg (1 µg) or better. It is used when the sample is only a few milligrams, for example a potent reference standard or a DSC sample.

  • Pharmaceutical
  • Biotech
  • Chemicals
  • Environmental testing
  • Academic research
  • Materials science
PREVENTIVE MAINTENANCE

Maintenance checklist

Daily / each use

  • Check levelling
  • Clean the pan and chamber
  • Perform and record the daily check with a small calibrated weight
  • Confirm the internal adjustment has run

Weekly

  • Clean the draught shield glass with a lint-free cloth
  • Check the ioniser works
  • Check the room temperature and humidity records for unusual drift

Monthly

  • Repeatability check with a small weight (or as defined in your SOP)
  • Inspect test weights and their box for dust or damage

Six-monthly / annual

  • Full calibration: repeatability, accuracy and minimum weight at the place of use
  • Recalibration of test weights by an accredited laboratory
  • Preventive maintenance by the manufacturer's engineer
SAFETY

Safety precautions

Before you work on the Microbalance
  • Weigh potent or toxic powders in a ventilated enclosure designed for microbalances, wearing gloves and, where required, a mask.
  • Clean spills straight away with the approved method; do not blow or brush powder into the air.
  • Handle weights and containers with forceps, never bare fingers.