A very sensitive balance that reads to 0.001 mg (1 µg) or better. It is used when the sample is only a few milligrams, for example a potent reference standard or a DSC sample.

  • Pharmaceutical
  • Biotech
  • Chemicals
  • Environmental testing
  • Academic research
  • Materials science
LEARN & PRACTICE

Test your Microbalance knowledge

Multiple-choice questions

Pick an answer to see whether you are right and why.

  1. Q1. A typical microbalance has a readability of:

    basic
  2. Q2. Why are containers placed on a microbalance with forceps instead of fingers?

    basic
  3. Q3. A dry powder in a plastic boat gives a reading that drifts steadily. The most likely cause is:

    intermediate
  4. Q4. The repeatability standard deviation of a microbalance is 0.8 µg. What is its minimum weight under USP <41>?

    intermediate
  5. Q5. Under USP <41>, if the standard deviation is smaller than 0.41 × d, you should:

    intermediate
  6. Q6. Why do many microbalances have their electronics in a separate unit?

    advanced
  7. Q7. Which test checks that a microbalance reads the true value of a calibrated weight?

    basic

Viva questions

Think of your answer first, then open it to compare.

  1. 1. When would you choose a microbalance instead of an analytical balance?

    When the sample is only a few milligrams, for example a scarce reference standard or a DSC sample. A 2 mg sample is far below the minimum weight of most analytical balances, so the error would be too large. A microbalance has a much lower minimum weight.

  2. 2. Why must the minimum weight be determined at the place where the balance is used?

    The minimum weight depends on repeatability, and repeatability depends on the local environment: draughts, vibration, temperature changes and static. A value measured in a quiet service lab would not be valid at a busy bench.

  3. 3. List the environmental requirements for a microbalance room.

    Stable temperature with slow changes only, moderate humidity (often around 45–60% RH) to reduce static, no direct air-conditioning draughts or sunlight, a free-standing heavy anti-vibration table, low foot traffic, and an ioniser for static. Exact values come from the manufacturer and your SOP.