An electrically heated, insulated cabinet that holds samples or glassware at a set temperature. In QC it is used mainly for loss on drying, drying of reference materials and drying of glassware.

  • Pharmaceutical
  • Food & beverage
  • Chemicals
  • Materials testing
  • Education
Photo of a Hot Air Oven
Photo: Karelj, Public domain (cropped), via Wikimedia Commons
TROUBLESHOOTING

Common problems, causes and fixes

Oven does not reach set temperature
Likely causes
  • • Failed heating element
  • • Door seal damaged or door left open
  • • Controller fault
What to do
  • • Have the element checked and replaced
  • • Replace the gasket and close the door properly
  • • Service the controller
Temperature overshoots or cycles widely
Likely causes
  • • Poorly tuned controller
  • • Sensor displaced or faulty
What to do
  • • Retune the PID controller
  • • Reposition or replace the sensor and recalibrate
Display reading differs from reference thermometer
Likely causes
  • • Sensor drift
  • • Controller offset not adjusted
What to do
  • • Recalibrate and apply the offset as per SOP
  • • Replace the sensor if drift is large
Poor uniformity (hot or cold spots)
Likely causes
  • • Fan not working
  • • Overloaded shelves
  • • Blocked vents
What to do
  • • Repair the fan
  • • Reduce load and space containers
  • • Clear the vents and repeat mapping
LOD results variable between analysts
Likely causes
  • • Different shelf positions
  • • Samples weighed while warm
  • • Inconsistent cooling time in the desiccator
What to do
  • • Use only mapped positions
  • • Cool to room temperature before weighing
  • • Fix the cooling time in the method
Over-temperature alarm or cut-out trips
Likely causes
  • • Controller failure
  • • Cut-out set too close to the working temperature
What to do
  • • Stop use and have the controller checked
  • • Set the cut-out as per manufacturer guidance