Fume hood

A ventilated enclosure that draws air away from the user to capture and remove harmful vapours, gases and dusts while working with chemicals.

  • Pharmaceutical
  • Chemicals
  • Universities and research
  • Environmental testing
  • Food testing
Photo of a Fume hood
Photo: Unknown, Public domain (cropped), via Wikimedia Commons
Follow your site SOP
This is a typical operating sequence for learning. Your instrument model and approved SOP take precedence.
STAGE 1

Before use

  1. 1Check the airflow monitor shows normal flow and the test label is in date.
  2. 2Clear the hood of unnecessary items and check nothing blocks the baffles.
  3. 3Set the sash to the marked working height or lower.
  • If the low-flow alarm sounds, do not start work.
STAGE 2

Working

  1. 1Work at least 15 cm (6 inches) inside the sash opening.
  2. 2Raise equipment on blocks so air can flow underneath.
  3. 3Keep your head outside the hood and move slowly to avoid disturbing the airflow.
  4. 4Close containers as soon as you have finished with them.
  • Do not use the hood as a store for chemicals or waste.
  • Avoid walking quickly past an open hood.
STAGE 3

After use

  1. 1Remove or close all containers.
  2. 2Wipe the work surface.
  3. 3Lower the sash fully.