TROUBLESHOOTING

Common problems, causes and fixes

Water check reads high
Likely causes
  • • Residue or deposit in the tube
  • • Bubble in the tube
  • • Water not pure or not degassed
What to do
  • • Clean the tube thoroughly and dry
  • • Refill slowly; check the camera
  • • Use fresh, freshly boiled and cooled purified water
Air reading does not match
Likely causes
  • • Tube not dry
  • • Solvent vapour still in the tube
What to do
  • • Dry longer with the air pump
  • • Rinse with a volatile solvent, then dry
Unstable readings
Likely causes
  • • Bubbles forming as the sample warms
  • • Temperature not stable
  • • Vibration from nearby equipment
What to do
  • • Degas the sample
  • • Wait for temperature stability
  • • Move the instrument away from vibration sources
Results higher than expected for a thick syrup
Likely causes
  • • Viscosity effect not corrected
What to do
  • • Turn on viscosity correction or use a method designed for viscous samples
Duplicates disagree
Likely causes
  • • Sample not mixed
  • • Particles settling in the tube
What to do
  • • Mix before filling
  • • Filter if the method allows, or use another technique