Incubator / BOD incubator
A temperature-controlled chamber that holds culture plates and tubes at a fixed temperature so microorganisms can grow. BOD incubators can heat and cool, so they also hold temperatures below room temperature, such as 20–25 °C for fungi.
- Pharmaceutical
- Biotech
- Food & beverage
- Water testing
- Environmental
- Research laboratories

TROUBLESHOOTING
Common problems, causes and fixes
Temperature above set point
Likely causes
- • Faulty sensor or controller
- • Room too hot for a heating-only unit
- • Refrigeration failure (BOD)
- • Heater relay stuck on
What to do
- • Check the reading with a calibrated thermometer
- • Move the unit or use a BOD incubator
- • Call service for the compressor or relay
- • Assess plates incubated during the excursion
Temperature below set point
Likely causes
- • Door left open or poor gasket seal
- • Heater failure
- • Ice build-up on the evaporator (BOD)
What to do
- • Close the door and replace a damaged gasket
- • Call service for the heater
- • Defrost and check the defrost cycle
Poor uniformity between shelves
Likely causes
- • Overloading or blocked air paths
- • Fan failure
- • Plates against the walls or fan
What to do
- • Reduce load and leave space between stacks
- • Check fan operation
- • Use only mapped positions
Display differs from reference thermometer
Likely causes
- • Sensor drift
- • Wrong calibration offset
What to do
- • Recalibrate and apply the correct offset
- • Replace the sensor if drift continues
Condensation inside plates or chamber
Likely causes
- • Plates not inverted
- • Large temperature difference between media and chamber
- • Water spills or humid samples
What to do
- • Incubate plates inverted
- • Pre-warm or dry plates before use
- • Clean and dry the chamber
Frequent alarms
Likely causes
- • Alarm limits set too tight
- • Frequent door opening
- • Monitoring probe in an unrepresentative position
What to do
- • Review alarm limits against the requirement
- • Plan access to reduce door openings
- • Relocate the probe to a mapped representative position
In LimsPulse · Calibration Management