Analytical balance · Calibration procedure
Eccentricity
Places the same weight in the centre and at four corners of the pan and checks the readings agree.

PURPOSE
Why this is checked
Analysts do not always place samples exactly in the centre. The eccentricity test shows how much the reading changes with position.
Frequency
Typically every 6 or 12 months, and after repair, relocation or a failed daily check, as defined in your calibration SOP.
Learn how the Analytical balance worksYOU WILL NEED
Standards & tools
- Calibrated test weight of about one third to one half of capacity, e.g. 100 g for a 220 g balance (as defined in your SOP)
BEFORE YOU START
Prerequisites
- Balance level, clean and warmed up
- Mark the positions on a diagram: centre, front-left, front-right, back-left, back-right
PROCEDURE
Step by step
- 1Zero the balance.
- 2Place the weight in the centre and record the reading.
- 3Move the weight to each corner position in turn (about halfway between centre and edge, or as the manufacturer describes) and record each reading.
- 4Return the weight to the centre and confirm the reading.
- 5Calculate the difference between each corner reading and the centre reading.
RECORD
Example observation table
| Position | Observed (g) | Difference from centre (g) |
|---|---|---|
| Centre | 100.0000 | 0.0000 |
| Front-left | 100.0002 | +0.0002 |
| Front-right | 99.9999 | −0.0001 |
| Back-left | 100.0001 | +0.0001 |
| Back-right | 100.0002 | +0.0002 |
CALCULATION
How to calculate the result
Eccentricity error = Corner reading − Centre reading (report the largest)
- Largest difference = 100.0002 − 100.0000 = 0.0002 g
- As a percentage: 0.0002 ÷ 100 × 100 = 0.0002%
ACCEPTANCE CRITERIA
Pass if
- Largest difference within the manufacturer's eccentricity specification or the limit in your SOP (for example not more than 0.0005 g at this load).
IF IT FAILS
What to do
- Check the pan is seated correctly and not touching the draught shield.
- Repeat the test. If it still fails, call service to adjust the weighing cell.
- Until repaired, the balance must not be used.