LimsPulse
Analytical balance · Calibration procedure

Daily performance check

A quick check before first use each day, weighing one or two calibrated weights to confirm the balance still reads correctly.

Photo of a Analytical balance
Photo: Karelj, Public domain (cropped), via Wikimedia Commons
PURPOSE

Why this is checked

A balance can be knocked, moved or affected by a temperature change overnight. The daily check catches a problem before it affects any weighing done that day.

Frequency

Every day before first use (and after the balance has been moved or cleaned).

Learn how the Analytical balance works
YOU WILL NEED

Standards & tools

  • Calibrated test weights, e.g. OIML class E2 or F1, with a valid certificate (commonly one small and one large weight, such as 20 g and 200 g)
BEFORE YOU START

Prerequisites

  • Balance switched on and warmed up
  • Balance level and pan clean
  • Internal adjustment performed if the balance has one
  • Weights at room temperature
PROCEDURE

Step by step

  1. 1Close the doors and zero the balance. Record that the display shows 0.0000 g.
  2. 2Using forceps, place the smaller test weight in the centre of the pan. Close the doors and record the stable reading.
  3. 3Remove the weight and confirm the reading returns to zero.
  4. 4Repeat with the larger test weight.
  5. 5Compare each reading with the certified value and record the result in the daily check log.
RECORD

Example observation table

Test weight (g)Observed (g)Difference (g)% errorLimitResult
0 (zero)0.00000.0000—0.0000 ± 0.0001 gPass
2020.0001+0.0001+0.0005%±0.05%Pass
200200.0003+0.0003+0.00015%±0.05%Pass
CALCULATION

How to calculate the result

% error = (Observed − Certified value) ÷ Certified value × 100
  1. 20 g weight reads 20.0001 g
  2. Difference = 20.0001 − 20.0000 = 0.0001 g
  3. % error = 0.0001 ÷ 20 × 100 = 0.0005%
ACCEPTANCE CRITERIA

Pass if

  • Each reading within the limit set in your SOP. Many labs use a tighter limit than the USP accuracy limit of 0.10% (for example ±0.05%) so that a drift is caught early.
  • Zero returns to 0.0000 g after removing the weight.
IF IT FAILS

What to do

  • Do not use the balance. Run the internal adjustment and repeat the check once.
  • If it fails again, label it 'Out of order' and inform the supervisor and the calibration team.
  • Assess weighings done since the last passing check.
MORE

Other Analytical balance calibrations

In LimsPulse · Calibration Management

LimsPulse schedules this calibration, records the result against limits, and blocks the instrument everywhere if it fails or goes overdue.

See how