Analytical balance · Calibration procedure
Accuracy
Weighs a calibrated test weight near the top of the range and compares the reading with its certified value. Also called a sensitivity test.

PURPOSE
Why this is checked
Confirms the balance's calibration (its sensitivity) is correct, so that readings across the range are true and not only precise.
Frequency
Typically every 6 or 12 months, and after repair, relocation or a failed daily check, as defined in your calibration SOP.
Learn how the Analytical balance worksYOU WILL NEED
Standards & tools
- Calibrated test weight between 5% and 100% of capacity, e.g. 200 g OIML class E2 for a 220 g balance, with certificate
BEFORE YOU START
Prerequisites
- Balance level, clean and warmed up
- Weight at room temperature and handled with forceps
PROCEDURE
Step by step
- 1Zero the balance with the doors closed.
- 2Place the test weight centrally and record the stable reading.
- 3Remove the weight and confirm the reading returns to zero.
- 4Repeat the measurement as many times as your SOP asks (often three) and use the mean.
- 5Compare the reading with the certified (conventional) mass of the weight.
RECORD
Example observation table
| Certified value (g) | Reading 1 (g) | Reading 2 (g) | Reading 3 (g) | Mean (g) | % error |
|---|---|---|---|---|---|
| 200.0000 | 200.0003 | 200.0004 | 200.0005 | 200.0004 | +0.0002% |
CALCULATION
How to calculate the result
% error = (Mean reading − Certified value) ÷ Certified value × 100
- Mean reading = (200.0003 + 200.0004 + 200.0005) ÷ 3 = 200.0004 g
- Difference = 0.0004 g
- % error = 0.0004 ÷ 200 × 100 = 0.0002%
ACCEPTANCE CRITERIA
Pass if
- Reading within 0.10% of the test weight value (USP <41>). Many SOPs also apply a tighter internal limit.
IF IT FAILS
What to do
- Run the balance's adjustment (internal or with an external weight) and repeat.
- If it still fails, label it 'Out of calibration' and call service.
- Assess the impact on results since the last passing calibration.