LimsPulse
APIs & raw materials · Active substance (powder or crystals)

Identification by UV

Supports the identity of a raw material by showing that its UV spectrum has absorption maxima (and minima) at the expected wavelengths and that its absorbance at the maximum agrees with the reference substance or with the stated specific absorbance.

  • UV-Vis spectrophotometry
  • UV-Vis
  • Analytical balance
  • Sonicator

Reference: General approach described in Ph. Eur. 2.2.25, USP <857> and USP <197>. The solvent, concentration, wavelengths and limits come from the individual monograph; always follow your approved specification and STP. The paracetamol figures here are an illustration only.

PRINCIPLE

Why this method works

A substance with a chromophore absorbs UV light at characteristic wavelengths. Under fixed conditions (solvent, pH, concentration) the positions of the maxima and the absorbance per unit concentration are characteristic of the substance. UV is less specific than IR, so it is usually one of several identification tests rather than the only one.

Instruments used

MATERIALS

Reagents & standards

ReagentGrade / specification
Reference substancePharmacopoeial reference standard or qualified working standard (if the monograph compares with a reference)
SolventAs stated in the monograph, e.g. 0.01 M sodium hydroxide for the paracetamol example, methanol or 0.1 M hydrochloric acid for others
WaterPurified water
SAMPLE PREPARATION

Preparing the sample

  1. 01

    Weigh

    Accurately weigh the sample (e.g. about 50 mg).

  2. 02

    Dissolve

    Dissolve in the specified solvent in a 100 mL flask; sonicate if needed.

  3. 03

    Dilute

    Dilute to the working concentration (e.g. 2 mL to 100 mL, about 10 µg/mL).

  4. 04

    Baseline

    Run a baseline with the solvent in matched quartz cells.

  5. 05

    Scan

    Record the spectrum, e.g. 220–350 nm.

  6. 06

    Evaluate

    Read λmax and absorbance; compare with the reference or specification.

Procedure

  1. 1Accurately weigh about 50 mg of the sample into a 100 mL volumetric flask.
  2. 2Add about 60 mL of the solvent (0.01 M sodium hydroxide in the paracetamol example), sonicate briefly to dissolve and make up to volume.
  3. 3Pipette 2.0 mL into a second 100 mL volumetric flask and make up to volume with the same solvent (about 10 µg/mL).
  4. 4Zero the instrument with the solvent and record the spectrum over the range stated in the monograph.
  5. 5Record the wavelengths of maxima and minima and the absorbance at the main maximum.
  6. 6If the result is to be expressed on the dried basis, use the loss on drying (or water content) value from the same batch.

Standard preparation

  1. 1Prepare the reference substance in the same way and at the same concentration as the sample, using the same solvent lot.
  2. 2Record its spectrum immediately before or after the sample under identical instrument settings.

Notes for analysts

  • The position of λmax often depends on pH and solvent: paracetamol, for example, absorbs at a different wavelength in acid than in alkali. Use exactly the solvent in the monograph.
  • Use quartz cuvettes for measurements below about 320 nm.
  • Prepare sample and reference in the same solvent lot to avoid small baseline differences.
  • Keep the absorbance in the reliable range by adjusting the dilution, but use the same dilution for sample and reference.
  • The paracetamol values here illustrate the calculation; check the current monograph for the official solvent, wavelength and limits.
TEST CONDITIONS

Method parameters (typical)

Wavelength rangeAs per monograph, e.g. 220–350 nm
Expected λmax (example)About 257 nm for paracetamol in 0.01 M sodium hydroxide
CellsMatched quartz cuvettes, 10 mm path length
BlankThe same solvent used to prepare the solutions
Spectral bandwidthAs per monograph or SOP (commonly 1–2 nm)
BEFORE YOU REPORT

System suitability & sequence

ParameterLimit
Instrument calibrationWavelength and absorbance accuracy within the current calibration
Absorbance rangeAbsorbance at λmax within the instrument's reliable range (typically about 0.2–1.0)
BaselineFlat baseline with the solvent blank

Typical sequence

  1. 1Solvent baseline
  2. 2Reference substance solution
  3. 3Sample solution
  4. 4Solvent check (baseline drift)
CALCULATION

Calculation with a worked example

A(1%, 1 cm) = A ÷ (c × b); % difference = (A(1%)sample − A(1%)reference) ÷ A(1%)reference × 100

Where

A
Absorbance measured at λmax
c
Concentration of the solution on the dried basis, in g/100 mL
b
Path length in cm (1 cm)
A(1%, 1 cm)
Specific absorbance: absorbance of a 1 g/100 mL solution in a 1 cm cell

Given

Sample weight
50.4 mg (loss on drying 0.2%) → 50.30 mg dried
Reference weight
50.1 mg (loss on drying 0.1%) → 50.05 mg dried
Dilution
100 mL, then 2 mL to 100 mL
Sample absorbance at 257 nm
0.720
Reference absorbance at 257 nm
0.716
Path length
1 cm

Working

  1. Sample concentration = 50.30 mg/100 mL × 2/100 = 1.006 mg/100 mL = 0.001006 g/100 mL
  2. Sample A(1%, 1 cm) = 0.720 ÷ (0.001006 × 1) = 715.7
  3. Reference concentration = 50.05 mg/100 mL × 2/100 = 1.001 mg/100 mL = 0.001001 g/100 mL
  4. Reference A(1%, 1 cm) = 0.716 ÷ (0.001001 × 1) = 715.3
  5. % difference = (715.7 − 715.3) ÷ 715.3 × 100 = +0.06%

λmax of sample and reference both at 257 nm; specific absorbance agrees within 0.06% (well inside a typical ±3% limit). The sample complies with this identification test.

ACCEPTANCE CRITERIA

What result passes

  • The sample spectrum shows maxima (and minima, if stated) at the same wavelengths as the reference substance, or at the wavelengths stated in the monograph, typically within ±1–2 nm or as the monograph states.
  • Where a reference substance is used: the absorbances (or absorptivities) at the maximum typically agree within 3%, as described in USP <197> for UV identification; use the limit in your specification.
  • Where the monograph gives a specific absorbance range at λmax, the calculated A(1%, 1 cm) lies within that range.
  • Where the monograph gives an absorbance ratio between two maxima, the calculated ratio lies within the stated range.
METHOD VALIDATION

Validation parameters (ICH Q2(R2))

ParameterTypical requirement
SpecificityShown by the shape and position of the maxima; UV alone is rarely specific, so it is combined with other identification tests (e.g. IR, HPLC retention time)
Compendial methodsVerified under actual conditions of use (solvent lot, instrument) rather than fully validated
In LimsPulse · Sample Management

Run this method in LimsPulse: specifications and STPs under version control, results calculated and checked against limits, and a signed Certificate of Analysis.

See how