LimsPulse
Drinking water · Drinking water, groundwater

Fluoride by Ion-Selective Electrode (with TISAB)

A fluoride-selective electrode develops a potential that depends on the logarithm of fluoride activity. Mixing each sample and standard 1 : 1 with TISAB fixes the ionic strength and pH and frees fluoride bound to aluminium or iron, so the meter reading can be converted to concentration using a calibration curve.

  • Environmental & water
  • Potentiometry with a lanthanum fluoride crystal electrode after 1 : 1 dilution with TISAB
  • pH meter
  • Analytical balance
  • Micropipette

Reference: Typical method based on APHA Standard Methods 4500-F⁻ C (ion-selective electrode method) and the IS 3025 part for fluoride. The pH meter is used in mV / ion mode. Follow your lab SOP.

PRINCIPLE

Why this method works

The sensing element is a lanthanum fluoride (LaF₃) single crystal. The potential across it follows the Nernst equation: E = E⁰ − S × log[F⁻], where S is about 59 mV per tenfold change at 25 °C. Total Ionic Strength Adjustment Buffer (TISAB) does three jobs: it sets a high, constant ionic strength so that activity is proportional to concentration; it buffers the pH at about 5–5.5, where neither HF formation (low pH) nor OH⁻ interference (high pH) matters; and it contains a complexing agent (such as CDTA) that releases fluoride from Al³⁺ and Fe³⁺ complexes.

Instruments used

MATERIALS

Reagents & standards

ReagentGrade / specification
TISABAcetate buffer pH about 5.0–5.5 with NaCl and CDTA (commercial TISAB II or prepared as per SOP)
Fluoride stockSodium fluoride (dried) in water, e.g. 100 mg/L F⁻; store in plastic
Working standardse.g. 0.2, 0.5, 1.0, 2.0 mg/L F⁻, bracketing the expected range
SAMPLE PREPARATION

Preparing the sample

  1. 01

    Collect

    Plastic (polyethylene) bottle; no preservative needed.

  2. 02

    Mix with TISAB

    25 mL sample + 25 mL TISAB.

  3. 03

    Standards

    Treat standards exactly the same way.

  4. 04

    Calibrate

    Read mV of standards, low to high; check slope.

  5. 05

    Measure

    Read sample mV at same temperature and stirring.

Procedure

  1. 1Collect samples in polyethylene bottles (fluoride can react with glass over time).
  2. 2Bring standards and samples to the same temperature (±1 °C).
  3. 3Pipette 25 mL of each standard and sample into plastic beakers and add 25 mL of TISAB to each. Mix.
  4. 4Set the meter to mV (or concentration) mode. Immerse the fluoride and reference electrodes in the lowest standard with gentle, constant magnetic stirring and record the stable reading. Repeat for each standard from low to high, rinsing and blotting the electrodes between solutions.
  5. 5Check that the slope is within the SOP range (typically −54 to −60 mV per decade at 25 °C).
  6. 6Measure each sample in the same way and read the concentration from the calibration. Re-read a mid standard every 10 samples or so.

Standard preparation

  1. 1Stock: dissolve 221.0 mg of dried NaF in water and make up to 1000 mL (1.00 mL = 100 µg F⁻). Store in a polyethylene bottle.
  2. 2Working standards: dilute the stock with water; each is then mixed 1 : 1 with TISAB like the samples.

Notes for analysts

  • In much of India, high fluoride in groundwater causes dental and skeletal fluorosis — fluoride is one of the most important health parameters for borewell sources.
  • Slow response or drifting readings usually point to a fouled crystal; polish or clean as per the manufacturer.
  • Never use glass beakers for low-level fluoride standards stored for long periods.
TEST CONDITIONS

Method parameters (typical)

ElectrodesLaF₃ fluoride ISE + reference electrode (or combination)
Sample : TISAB1 : 1 by volume
TemperatureStandards and samples within ±1 °C of each other
StirringSame speed for all solutions
CalibrationAt least three standards bracketing the samples
BEFORE YOU REPORT

System suitability & sequence

ParameterLimit
Electrode slopeTypically −54 to −60 mV/decade at 25 °C (or as per SOP)
Check standardMid standard re-read within ±5–10% of nominal (as per SOP)
Spike recoveryTypically 90–110%

Typical sequence

  1. 1Standards low → high
  2. 2Check slope
  3. 3Samples
  4. 4Mid-standard drift check every ~10 samples
  5. 5Spiked sample
CALCULATION

Calculation with a worked example

Formula
E = E₁ + S × log(C ÷ C₁) → C = C₁ × 10^((E − E₁) ÷ S)

Where

E
Sample potential (mV)
E₁
Potential of a reference standard of concentration C₁ (mV)
S
Electrode slope (mV per decade; negative for an anion electrode)
C
Fluoride concentration of the sample (mg/L)

Given

Standards (mg/L F⁻) and mV
0.5 → +17.6; 1.0 → 0.0; 2.0 → −17.6
Slope S
−58.5 mV/decade (17.6 mV per doubling = 58.5 × log 2)
Sample reading
+5.2 mV

Working

  1. 1

    Take C₁ = 1.0 mg/L, E₁ = 0.0 mV

  2. 2
    (E − E₁) ÷ S = (5.2 − 0.0) ÷ (−58.5) = −0.0889
  3. 3
    C = 1.0 × 10^(−0.0889) = 0.815 mg/L
  4. 4

    Samples and standards were both diluted 1 : 1 with TISAB, so no further dilution factor is needed

Fluoride = 0.81 mg/L (below the IS 10500 acceptable limit of 1.0 mg/L)

ACCEPTANCE CRITERIA

What result passes

  • As per IS 10500:2012: acceptable limit 1.0 mg/L, permissible limit 1.5 mg/L in the absence of an alternative source; check the current edition.
  • WHO guideline value: 1.5 mg/L; national authorities may set lower values considering climate and water intake.
METHOD VALIDATION

Validation parameters (ICH Q2(R2))

ParameterTypical requirement
LinearityNernstian response across the calibration range
AccuracySpike recovery and comparison with a certified reference water
InterferencesHigh aluminium may exceed the complexing capacity of TISAB; check with spikes
In LimsPulse · Sample Management

Run this method in LimsPulse: specifications and STPs under version control, results calculated and checked against limits, and a signed Certificate of Analysis.

See how