LimsPulse
General lab & storage

Water bath

A tank of water held at a constant set temperature, used to warm, incubate or melt samples gently and evenly.

  • Pharmaceutical
  • Microbiology
  • Food & dairy
  • Clinical laboratories
  • Research
Photo of a Water bath
Photo: Athikhun.suw, CC BY-SA 4.0 (cropped), via Wikimedia Commons
PREVENTIVE MAINTENANCE

Maintenance checklist

Daily / each use

  • Check the water level and top up with purified water
  • Check the displayed temperature

Weekly

  • Drain, clean and refill with fresh purified water
  • Wipe the lid and racks

Monthly / quarterly

  • Descale if deposits build up
  • Check the over-temperature cut-out

Six-monthly / annual

  • Calibrate temperature accuracy and uniformity against a reference thermometer
SAFETY

Safety precautions

Before you work on the Water bath
  • Hot water and containers can cause burns; use heat-resistant gloves.
  • Keep electrical connections dry and unplug the bath before draining or filling it.
  • Do not heat flammable liquids in an open water bath.
In LimsPulse · Calibration Management

Keep every Water bath in control: usage logs, calibration and maintenance schedules, and automatic blocking of overdue instruments.

See how