Tap Density Tester

Taps a graduated cylinder of powder a set number of times from a fixed height so the bed packs down. The starting and final volumes give bulk density, tapped density, Carr's index and Hausner ratio.

  • Pharmaceutical
  • Nutraceutical
  • Chemicals
  • Food & beverage
  • Powder metallurgy
Illustration of Tap Density Tester
TROUBLESHOOTING

Common problems, causes and fixes

Poor repeatability between duplicates
Likely causes
  • • Different filling technique
  • • Sample not sieved or not representative
  • • Uneven powder surface
What to do
  • • Standardise pouring; train analysts
  • • Sieve and mix before sampling
  • • Read at the mean level of the surface
Volume keeps decreasing after many taps
Likely causes
  • • Very cohesive or fragile powder
  • • Particles breaking under tapping
What to do
  • • Continue tapping in increments as the method allows
  • • Report the number of taps used
Tap count wrong
Likely causes
  • • Faulty counter or sensor
  • • Belt slipping
What to do
  • • Verify by manual count; service the counter
  • • Replace the belt
Drop height out of tolerance
Likely causes
  • • Worn cam or anvil
  • • Loose holder
What to do
  • • Replace worn parts and recalibrate
  • • Tighten fixings