Tap Density Tester
Taps a graduated cylinder of powder a set number of times from a fixed height so the bed packs down. The starting and final volumes give bulk density, tapped density, Carr's index and Hausner ratio.
- Pharmaceutical
- Nutraceutical
- Chemicals
- Food & beverage
- Powder metallurgy
Follow your site SOP
This is a typical operating sequence for learning. Your instrument model and approved SOP take precedence.
STAGE 1
Preparation
- 1Confirm the tester is within its calibration due date (drop height and tap rate).
- 2Pass the sample through the stated sieve to break up loose lumps.
- 3Weigh the stated amount (commonly about 100 g for a 250 mL cylinder) to 0.1% accuracy.
- Do not tap or knock the cylinder while filling; that already compacts the powder.
STAGE 2
Bulk volume
- 1Tilt the cylinder and pour the powder in gently.
- 2Level the surface carefully without compacting it.
- 3Read the unsettled volume V0 to the nearest graduation.
STAGE 3
Tapping
- 1Fix the cylinder in the holder.
- 2Run the tap counts stated in the method (for example 10, 500 and 1250 taps) and read the volume after each.
- 3If the difference between successive readings is larger than the method limit, continue tapping in further increments until it is within the limit.
- 4Record the final tapped volume Vf.
- Keep fingers clear of the moving holder during tapping.
STAGE 4
After use
- 1Empty and clean the cylinder
- 2Clean powder from the base and holder
- 3Complete the usage log