Tap Density Tester

Taps a graduated cylinder of powder a set number of times from a fixed height so the bed packs down. The starting and final volumes give bulk density, tapped density, Carr's index and Hausner ratio.

  • Pharmaceutical
  • Nutraceutical
  • Chemicals
  • Food & beverage
  • Powder metallurgy
Illustration of Tap Density Tester
WORKING PRINCIPLE

How Tap density tester works

Repeated mechanical tapping rearranges particles and removes air spaces, so the bed volume falls to a near-constant tapped volume. Density is mass divided by volume, and the change between bulk and tapped volume reflects inter-particle friction and cohesion.

FLOW

The path of a sample

  1. 01

    Sieve and weigh

    The powder is passed through a sieve to break up lumps and an accurately weighed amount is taken.

  2. 02

    Fill

    The powder is poured gently into a dry graduated cylinder without compacting it.

  3. 03

    Read V0

    The unsettled (bulk) volume is read to the nearest graduation.

  4. 04

    Tap

    The tester drops the cylinder a fixed height a set number of times.

  5. 05

    Read Vf

    Volumes are read after set tap counts until the change is within the limit; the final reading is the tapped volume.

  6. 06

    Calculate

    Bulk density, tapped density, Carr's index and Hausner ratio are calculated.

COMPONENTS

Main parts and what they do

1

Graduated cylinder

Typically 250 mL, readable to 2 mL, of stated mass; 100 mL cylinders are used for small samples.

2

Cylinder holder and anvil

Holds the cylinder and lets it fall onto a fixed base.

3

Cam drive

Lifts and releases the cylinder to give a fixed drop height at a set rate.

4

Tap counter / controller

Sets and counts the number of taps and stops automatically.

5

Sieve

Breaks up agglomerates before filling (often about 1 mm aperture).

KEY CONCEPTS

Terms every analyst should know

Bulk density

Mass ÷ untapped volume (g/mL). Includes the air spaces between particles.

Tapped density

Mass ÷ final tapped volume (g/mL). Always equal to or higher than bulk density.

Carr's index (compressibility index)

100 × (V0 − Vf) ÷ V0. Higher values mean poorer flow.

Hausner ratio

V0 ÷ Vf. Values close to 1 mean good flow.

Drop height and tap rate

Fixed by the pharmacopoeial method; changing them changes how fast and how far the powder settles.