Moisture Analyser (Halogen / IR LOD Balance)

A balance with a built-in halogen or infrared heater. It dries a sample on the pan and calculates the loss on drying automatically, usually in a few minutes.

  • Pharmaceutical
  • Food
  • Plastics
  • Chemicals
  • Agriculture
Photo of a Moisture Analyser (Halogen / IR LOD Balance)
Photo: Nick Birse, CC BY-SA 4.0 (cropped), via Wikimedia Commons
TROUBLESHOOTING

Common problems, causes and fixes

Results lower than oven LOD
Likely causes
  • • Temperature too low
  • • Switch-off criterion too loose
  • • Sample too thick
What to do
  • • Re-validate method settings
  • • Tighten the criterion
  • • Use a thinner, even layer
Sample burns or turns brown
Likely causes
  • • Temperature too high
  • • Sample sensitive to heat
What to do
  • • Lower the temperature and lengthen the time
  • • Use a glass fibre filter over the sample
Unstable reading
Likely causes
  • • Draughts or vibration
  • • Instrument not level
What to do
  • • Move to a stable location
  • • Level the instrument
Poor repeatability
Likely causes
  • • Variable sample mass or spreading
  • • Non-uniform sample
What to do
  • • Use the same mass and spreading each time
  • • Mix the sample before sampling